Apogee Instruments MQ-210 Owner's Manual - Page 13

Browse online or download pdf Owner's Manual for Measuring Instruments Apogee Instruments MQ-210. Apogee Instruments MQ-210 18 pages. Underwater quantum meter

Apogee Instruments MQ-210 Owner's Manual
13
When a quantum sensor that was calibrated in air is used to make underwater measurements, the sensor reads low. This
phenomenon is called the immersion effect and happens because the refractive index of water (1.33) is greater than air
(1.00). The higher refractive index of water causes more light to be backscattered (or reflected) out of the sensor in water
than in air (Smith,1969; Tyler and Smith,1970). As more light is reflected, less light is transmitted through the diffuser to the
detector, which causes the sensor to read low. Without correcting for this effect, underwater measurements are only relative,
which makes it difficult to compare light in different environments.
The MQ-210 sensor has an immersion effect correction factor of 1.08. The immersion effect correction factor is already
accounted for in the MQ-210 meter firmware so there is no need to apply the correction factor to your measurements. If you
wish to use your meter to make measurements in air, simply divide the measured number by the immersion effect (1.08).
When making underwater measurements, only the sensor and cable can go in the water. The handheld meter is not
waterproof and must not get wet. If the meter might get wet from splashing, we recommend placing it in a plastic bag or
other container to help protect it from accidentally getting wet.
Further information on underwater measurements and the immersion effect can be found at
http://www.apogeeinstruments.com/underwater-par-measurements/.