DataApex Shimadzu LC-10 Manual - Page 13
Browse online or download pdf Manual for Water Pump DataApex Shimadzu LC-10. DataApex Shimadzu LC-10 19 pages. Metal free pump
Shimadzu LC-10/20 Pump
Idle State
An item specifying the overall flow rate through the column outside the
instrument method. The following options are possible:
Pump Off
The flow rates of all components are zero.
Caution:
Be careful as this setting may damage the column in some cases.
Initial
The flow rate is defined by the first row of the gradient table (the Initial
row).
Standby
The flow rate is the same as in the STANDBY mode and, accordingly,
corresponds to the value entered in Standby Flow field.
The IDLE state comes into effect each time an Instrument is opened, at the
end or after abortion of an analysis by the Abort command, and is also
maintained after the Clarity program is shut down.
The mixing ratio of individual components in both the IDLE and STANDBY
states is given by the first row of the Gradient Table (the Initial row).
Note:
There is a step change in the flow and components percentage from the
values specified for the STANDBY state to those specified for the IDLE
state if the Idle State field is not set to Standby.
4 Using the control module
- 9 -