Active Motif 53080 Manual - Page 11
Browse online or download pdf Manual for Medical Equipment Active Motif 53080. Active Motif 53080 15 pages. Cooled sonication platform
Using the Cooled Sonication Platform
Benchtop Tube Coolers
The Benchtop Tube Coolers can keep the temperature of a sample at or below 4°C for up to
2 hours of sonication if the cooler has been placed in a -20°C freezer overnight. If a large amount
of time will pass between each sonication, return the cooler be to a -20°C freezer between
samples. Do not place coolers at -80°C for more than 30 minutes as they may be damaged.
Note:
If the Tube Cooler was kept in a -20°C freezer overnight, it will be sufficiently
cold to freeze your samples in about 5 minutes, unless you begin sonication.
Therefore, to prevent your sample from freezing, do not place it in the cooler
until you are ready to begin sonication. Keep your sample on ice and the cooler
in the freezer until all preparations for sonication have been completed.
Sonication
Ensure that the converter/probe is positioned properly in the converter/probe bracket and that
the platform settings are those that you determined to be appropriate for your tube type and
sample volume. If need be, reposition the converter/probe to the mark you placed on it and
re-zero the platform, then re-set it to its desired position.
Turn on your probe sonicator and adjust its settings to your preference. Press the latch and move
the platform down the rod. Place an open tube containing your sample in the Tube Cooler, then
place the cooler in the platform stand. Raise the platform until it is against the Stop Collar, and
ensure that the appropriate counter setting has been reached. Lift the tube up and move it around
slightly to verify that all is as expected. After verifying this, ensure that the tube is pressed down
all the way in the cooler. You are now ready to begin sonication following your protocol.
After sonication is complete, press the latch and lower the platform so that you can remove the
sample and place it on ice. Process additional samples, or clean the sonicator probe per your lab' s
standard practices, and return the Tube Cooler to the -20°C freezer.
www.activemotif.com
8