Active Motif 53080 Manual - Page 13
Browse online or download pdf Manual for Medical Equipment Active Motif 53080. Active Motif 53080 15 pages. Cooled sonication platform
Loosen the two set screws on the front of the
bracket, then loosen the small set screw on top of
the bracket. You can then slide out the bolt and
bobbin around which the continuous force spring
is wound (top image). Be sure to hold the spring as-
sembly firmly to prevent it from unwinding from the
bobbin, which could damage it. Take care with the
metal edges of the spring, as they could potentially
cut your finger.
While you are holding the spring/bobbin/bolt, slide
the converter/probe bracket off the top of the rod
and replace it with the alternate spring bracket. You
can put the spring assembly into the new bracket
(middle image).
Use one of the two large set screws that were in
the converter/probe bracket to secure the alternate
spring bracket to the rod. Likewise, use the small set
screw that was in the converter/probe bracket to
secure the spring in the alternate spring bracket.
You can then adjust the height of the bracket to
where you would like it to be and firmly secure the
set screw. If you will need to secure some other kind
of bracket to hold an oversized converter/probe,
you will need to place the bracket as low as it will
go on the rod (lower image).
To reassemble the platform back inside the sound
enclosure, angle the top of the rod into the probe
chimney. Gently slide the bottom of the rod
through the hole in the bottom of the sound enclo-
sure. Align the top of the rod into the designated
hole at the top of the sound enclosure. Make sure
the platform faces the interior of the box with
enough clearance for the door to close. Secure the
rod in place by tightening the screw on the top of
the sound enclosure and reassembling the three
screws and bracket on the bottom of the sound
enclosure. Set the EpiShear Cooled Sonication
Platform upright and proceed to Calibration of the
Platform Counter on page 6.
www.activemotif.com
10