iHealth ABP100 User Manual - Page 11

Browse online or download pdf User Manual for Blood Pressure Monitor iHealth ABP100. iHealth ABP100 16 pages. Ambulatory blood pressure monitor

iHealth ABP100 User Manual
LED display reads
"Er1"
LED display reads
"Er2"
LED display reads
"Er3"
LED display reads
"Er4"
LED display reads
"Er5"
LED display reads
"Er6"
LED display reads
"Er7"
LED display reads
"Er8"
LED display reads
"ErA"
LED display reads
" Er"
LED display reads
an abnormal
result
Bluetooth
connection
unstable
No response when
you press button
Fail to detect systolic pressure
Fail to detect diastolic pressure
Pneumatic system blocked or cuff
is too tight during inflation
Pneumatic system leakage or cuff
is too loose during inflation
Cuff pressure above 300mmHg
More than 160 seconds with cuff
pressure above 15 mmHg
memory accessing error
Device parameter checking error
Pressure sensor parameter error
Bluetooth communicate error
The cuff position was not correct
or it was not
properly tightened
Body posture was not correct
during testing
Speaking, moving arm or body,
being angry,
excited or nervous during test
Bluetooth
unsuccessful, monitor is
abnormal,
or
electromagnetic interference
is present
Incorrect
operation
electromagneticinterference
Apply the cuff correctly and try
again
Measure
minutes. If the monitor is still
abnormal, please contact the
local distributor or the factory.
Reset
the
about 10 seconds, then connect
the mobile device correctly and
try again, If the monitor is still
abnormal, please contact the
local distributor or the factory.
Review
instructions and retest
Review
instructions and retest
Retest
speaking or movement during
the test
Reset iOS device. Reset monitor
connection
by pressing the
holding for about 10 seconds.
strong
Make sure the monitor and iOS
device are away from other
electrical equipment. Please see
Warning
or
strong
Press the
11 / 16 
again
after
five
monitor
by
pressing
button and holding for
the
cuff
application
body
posture
when
calm;
avoid
button and
button and hold