AUTOSCALER AW30C Manual - Page 7

Browse online or download pdf Manual for Media Converter AUTOSCALER AW30C. AUTOSCALER AW30C 14 pages.

FEATURES

The AUTOSCALER® is designed for Dental Prophylaxis Treatment when
used with the appropriate Cavitron® brand – Series: "P", "SLIMLINE"™,
"FSI"™ or "TFI"™ 25,000Hz series inserts*. If you have purchased the
AUTOSCALER® 30,000Hz model you must use the 30,000Hz frequency
inserts. Inserts are not interchangeable. The AUTOSCALER® is acceptable
for use in scaling and cleaning surfaces of teeth in conjunction with other
prophylactic measures.
This Build-In style AUTOSCALER® unit is NOT intended for user/owner
install. It requires a technician familiar with dental operatory Air and
Water circuitry. This product needs installation of the following for
proper operation:
1. The Handpiece Assembly. Routing of 18 gauge wires through the
dental operatory to the AUTOSCALER® console.
2. Water delivery/control for the unit. The AUTOSCALER® does not come
equipped with any water control devices/controllers, however proper
water flow is mandatory for the operation of this product. See oper-
ating instructions later in this booklet for proper operating parameters.
3. Power Control Pot. Routing of (2) 18 gauge wires through the dental
operatory to the AUTOSCALER® console.
4. An independent Air Source to activate the AUTOSCALER® supplied by
the dental operatory (i.e. independent air/water block with auto
holder). The air supply must relieve to zero pressure after each
depression of the main air rheostat.
AUTOSCALER® is a trade mark of South East Instruments, LLC.
Recommendations by the manufacturer contained within this instruction
booklet are suggestions only. Other options may be available or mandated
by the professional or regulating agencies.
The source of the air signal, water supply, and installation are not the
responsibility of South East Instruments, LLC. Any damage to the
AUTOSCALER®, PATIENTS or DENTAL OPERATORY resulting from improper
installation or configuration is the sole responsibility of the installing dealer
or technician. The AUTOSCALER® BUILD-IN Models AW25BI or AW30BI are
sold as component devices that require further installation by a party
appointed by the end user. If the installing technician or end user is in doubt
as to correct installation procedures they should stop immediately and call
South East Instruments, LLC for further instructions at (800) 648-9445 before
attempting to proceed with the installation.
*Cavitron®, SLIMLINE™, FSI™, and TFI™ are registered trade marks of Dentsply Sirona
13

BUILD-IN MODEL INSTALLATION GUIDE

This GUIDE is to assist the technician. Most dental chairs utilize similar
configurations. It is the responsibility of the technician to properly
install and confirm proper function with the particular chair
manufacturer configuration.
DO NOT attempt to install this device without an independent air/
water block with automatic handpiece holder. The technician must
be familiar with dental chair air water logic to correctly install this
device. Most installations convert the existing 3
high or low speed drill block for this purpose.
There MUST be (4) 18
GAUGE wires running from
the junction box,
THROUGH THE ARM to the
tray for handpiece and power
control connections.
The AUTOSCALER® (CONSOLE) is mounted under the chair or
(JUNCTION BOX). NOTE: If the AUTOSCALER® receives an air signal
from another source (i.e. high speed or low speed drill) it can damage
the device and the handpiece assembly.
WATER FLOW to the handpiece is received from the air/water block (NEEDLE
VALVE). When the main air rheostat is pressed, an air signal is sent to the
AUTOSCALER® console to (cavitate the insert) and the air/water block opens to
release water flow to the handpiece. Quickly adjust the water knob to get
water flowing.
rd
th
or 4
air/water
The handpiece and
(POWER CONTROL)
are mounted on the (tray).
An (AUTOMATIC
HOLDER) must be used
for the Autoscaler®
handpiece.
14