AGS Merlin 1000S Installation & Operation Manual - Page 5

Browse online or download pdf Installation & Operation Manual for Laboratory Equipment AGS Merlin 1000S. AGS Merlin 1000S 12 pages. Gas utility isolation controller
Also for AGS Merlin 1000S: User & Installation Manual (10 pages), Installation, Operation And Maintenance Manual (10 pages), Installation & Operation Manual (16 pages)

AGS Merlin 1000S Installation & Operation Manual
Installation & Operation Manual

Typical Installation Arrangement

Terminal Connections

1. POWER / LINE IN
110-120V AC Power should be supplied to the [POWER / LINE IN] terminal and fused at 3A.
2. GAS SOLENOID VALVE OUTPUT
110-120V AC electrical power supplied from the [VALVE OUT] connector using a 3 core cable can be
connected to a gas solenoid valve which can shut the gas supply on alarm status.
Pressure sensors should be screwed to the downstream port of the valve.
3. BMS OUTPUT
Connections are available on the board for Building Management Systems.
[NO Normally Open] [COM Common] [NC Normally Closed] These are volt free connections.
This is a relay that changes state in alarm or when the gas is on/off and can be used in conjunction with
the 12V DC output and other external relays that affect other devices and controls such as purge fans
and audible alarms etc. See section; Switch Settings - for BMS options
Rev: 10
02-21
Merlin 1000S