V S A Q u i c k S t a r t G u i d e
1. Getting Started—
• Place the AquaLab VSA on a flat surface. Use the bubble
level(Figure 1: Adjustment and bubble level) and use the
adjustable feet to level the instrument. The small rubber
feet on the back of the lower plate are to prevent tip-over
and do not need to be touching the table.
• Use the AquaLab VSA where environmental temperature
and humidity remain fairly stable.
Adjustment feet:
Two on front and one
on the middle back.
Figure 1: Adjustment and bubble level.
• The AquaLab VSA needs to be on a stable platform and
evaluated to verify shipping did not damage the scale
performance. To test scale stability:
1. Turn on the VSA and allow instrument to warm
up for 2 hours.
2. Navigate to the Diagnostics screen in the
Configuration menu and press .
3. Scroll down to the screen that shows the weight.
Readings should vary no more than 0.5 mg.
Bubble Level
Lower Block