V S A Q u i c k S t a r t G u i d e

1. Getting Started—

  •  Place the AquaLab VSA on a flat surface. Use the bubble   
       level(Figure 1: Adjustment and bubble level) and use the
    adjustable feet to level the instrument. The small rubber 
    feet on the back of the lower plate are to prevent tip-over
    and do not need to be touching the table.
  •  Use the AquaLab VSA where environmental temperature 
       and humidity remain fairly stable.
Adjustment feet: 
Two on front and one
on the middle back.
Figure 1: Adjustment and bubble level.
  •  The AquaLab VSA needs to be on a stable platform and 
    evaluated to verify shipping did not damage the scale
    performance. To test scale stability:
      1.  Turn on the VSA and allow instrument to warm 
    up for 2 hours.
      2.  Navigate to the Diagnostics screen in the
         Configuration menu and press  .
      3.  Scroll down to the screen that shows the weight. 
         Readings should vary no more than 0.5 mg.
Bubble Level
Lower Block