AFRISO DIM 20 Instruction Manual - Page 17

Browse online or download pdf Instruction Manual for Measuring Instruments AFRISO DIM 20. AFRISO DIM 20 20 pages. Digital pressure gauge

AFRISO DIM 20 Instruction Manual
CAL EP
LD FAC
OK
ERROR
DIM 20
Menu 9: Calibrating the maximum end point
Should you notice a deviation of the indicated value in relation to the
maximum end point, the unit can be recalibrated. A suitable refer-
ence pressure is necessary for this task.
1.
Call up menu point "CAL EP".
2.
Press OK.
"CAL EP" starts to blink.
3.
Pressurize the instrument with a reference pressure. This pres-
sure must correspond to the pressure of the maximum end
point.
4.
Press
to read the pressure.
5.
Press OK to memorize the currently displayed signal as the
maximum end point.
Should this configuration lead to a worsening of the original results,
for instance due to lack of a suitable reference pressure, the original
factory calibration as per original order can be re-instated via the
menu 10 "LD FAC".
Menu 10: Loading of factory calibration
1.
Press OK.
2.
Press
to load the factory calibration according to the original
order.
After the loading process, the message "LOADED" appears
briefly on the display followed by "OK".
The unit then exits this menu point automatically.
Checking of calibrated values
Upon exiting the menu the adjusted parameters are checked in rela-
tion to each other and with reference to the basic instrument data. If
the message "OK" appears in the left hand bottom of the display, the
configuration was successful. If the message "ERROR" appears, then
at least one of the adjusted values is outside the permissible range.
This is for instance the case when the Digital Pressure Gauge has a
nominal pressure range of 400 bar and in the menu "DECIMA" four
decimal points were programmed. If the error message appears, the
last adjusted and functionally correct parameters are retained. Dur-
ing the configuration of the unit a conversion of the measuring
ranges (Menus "UPPER" and "LOWER") to the new units takes place
only after exiting the menu system..
Depending on the pressure range and the number of displayable fig-
ures, it may not be possible to use all measurement units.
Operation
17