Deckma OMD Series Servicing And Cleaning Manual - Page 13
Browse online or download pdf Servicing And Cleaning Manual for Measuring Instruments Deckma OMD Series. Deckma OMD Series 13 pages. Cleaning procedure of measuring cell mz-09
DECKMA
HAMBURG
1.7 Desiccator Check
✓
Figure 20
X
Figure 21
1. Press "SYS" Button, select Measuring
Cell, confirm.
2. Internal data will be displayed.
3. The relative humidity [%rH] should be
below 10%rH.
4. The dew-point Temperature should be
below 0° C.
5. If not, exchange desiccator.
6. The desiccator also has to be exchanged,
if
humidity
Do not expose the Desiccator to ambient
air. Do not leave the Desiccator opening
uncovered.
Page: 13
is
higher
40%rH.