Deckma OMD Series 서비스 및 청소 매뉴얼 - 페이지 13

{카테고리_이름} Deckma OMD Series에 대한 서비스 및 청소 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Deckma OMD Series 13 페이지. Cleaning procedure of measuring cell mz-09

DECKMA
HAMBURG

1.7 Desiccator Check

Figure 20
X
Figure 21
1. Press "SYS" Button, select Measuring
Cell, confirm.
2. Internal data will be displayed.
3. The relative humidity [%rH] should be
below 10%rH.
4. The dew-point Temperature should be
below 0° C.
5. If not, exchange desiccator.
6. The desiccator also has to be exchanged,
if
humidity
Do not expose the Desiccator to ambient
air. Do not leave the Desiccator opening
uncovered.
Page: 13
is
higher
40%rH.