CORNING LSE 6710 Instruction Manual - Page 4

Browse online or download pdf Instruction Manual for Laboratory Equipment CORNING LSE 6710. CORNING LSE 6710 9 pages. Digital vortexer
Also for CORNING LSE 6710: Instruction Manual (12 pages)

3.4 Environmental Requirements

The Corning® LSE™ Multipurpose Digital Vortexer has been built for operating in both
cold and heated laboratory environments and in high humidity environments as follows:
At temperatures between 0°C and +65°C
◗ ◗
In humidity up to 85% (without condensation)
◗ ◗

4.0 Installation of Vortexer Platforms

Cat. No.
Platform Descriptions
48010
Platform for 4 microplates
48011
Platform with non-slip rubber mat
48012
Spring loaded platform for flasks, bottles, or tube racks
The Corning LSE Multipurpose Digital Vortexer requires the installation of the selected
platform. Please follow the instructions below for installing the platform.
Sliding Stabilizers
Before you attach the platform, make sure that the slide stabilizers are properly
fastened, adjusted, and lubricated with the included grease.
The flat mat platform has a circle cut-out in the rubber and screw holes in the center.
Remove the cut-out rubber circle and screw down the 4 screws onto the Vortexer. After
attaching the platform, remove the non-adhesive backing paper on the rubber circle,
and paste it back into place.
The grease is used to lubricate the slide stabilizers. It is sufficient for 1,000 working
hours (approx. 1 year). After that time you should remove the platform and clean,
adjust, and re-grease the slide stabilizers. If they are worn out, you should replace them
with new stabilizers and set to appropriate height.
6
C ORNING LS E MU LTI PU R PO S E D I G I TA L VO RTEXER
Grease

5.0 Operating Instructions

5.1 Introduction

1
Run
Time
Power
Speed (RPMx10) time (min)
1. POWER switch– switches Vortexer ON (light on) or OFF.
2. RUN green signal light – lit when the Vortexer is running.
3. TIME yellow signal light – lit when the Vortexer is set on time.
4. RPM yellow signal light – lit when the Vortexer is set on rpm.
5. ENCODER – by rotating the encoder right (+) or left (-) you can change the set TIME or
RPM. Push the encoder to switch between TIME and RPM set values. If you rotate the
ENCODER knob fast, the values go up or down on the display very quickly.
6. START/STOP button – press to START or STOP operation.

5.2 Basic Operation

Run
Time
Power
Speed (RPMx10) time (min)
Press the POWER switch on the control panel. The Vortexer automatically detects
◗ ◗
supply frequency of 50 or 60Hz, and displays F50 or F60. Then the display switches to
show the set TIME and illuminates the TIME light.
2
3
4
5
Speed
Start
ENCODER
Stop
PUSH
(Time/Speed)
Speed
Start
ENCODER
Stop
PUSH
(Time/Speed)
Instruction Manual
6
7