Avantes AvaLight-Hal-CAL-Mini Manual - Page 6

Browse online or download pdf Manual for Lighting Equipment Avantes AvaLight-Hal-CAL-Mini. Avantes AvaLight-Hal-CAL-Mini 11 pages.

Avantes AvaLight-Hal-CAL-Mini Manual
CC connection
Position to insert CC-VIS/NIR of the system to be calibrated. CC can be locked with an allen screw.
I/O connector
For automatic disable control to be controlled by software it is needed to connect an AvaSpec
spectrometer to the AvaLight-HAL-CAL-Mini with the interface cable IC-DB26-2.
The signals on the I/O connector are described below.
Pin configuration
Pin
Name
8
Vout
Output
10
GND
11
Disable
Input
Below the electrical characteristics for the input signals.
Electrical characteristics
Name
Disable

2.2 12VDC Power plug

The power supply specifications are mentioned in the table below
Name
Voltage
Power

2.3 Settings for disable

With the disable function that is present on the I/O connector the user can disable the optical
output of the light source by setting this signal to high-level.
When the light source is used in combination with an AvaSpec spectrometer, interface cable IC-
DB26-AvaSphere-0.5 and AvaSoft, the output can be disabled or enabled by AvaSoft.
See appendix A.
Revision: 1-00
© Avantes BV
Description
Nominal 12VDC output
Ground
Input, used to disable/shut down the output
Low level detection
< 1.0V
Minimum
10V
18Watt
www.avantes.com
High level detection
> 2.0V
Typical
12V
Maximum voltage
13.2V
Maximum
15V
5