Aliyiqi HF-500 Quick Start Manual - Page 9
Browse online or download pdf Quick Start Manual for Test Equipment Aliyiqi HF-500. Aliyiqi HF-500 15 pages. Hf digital display push-pull force gauge
Seven, boot display
Eight, function introduction
1. Setting function
a. Mode switching: After the boot is complete, press \"Settings \" key, \"TEST \" appears, press \"Settings\" key
again, 0001 mode appears, use
\"Keys can choose different modes. As shown in the figure below:
0001 Standard test mode (factory default mode): In this mode, three states can be set, namely real-time load
state, peak hold state and automatic peak state. When there is no \"PEAK \" on the display, it is the real-time
value state of the load, and the test value changes with the change of the load; press \"Peak \" key, when the
word \"PEAK \" is displayed, it is the peak hold state The displayed test value is the MAX value of the test
(regardless of tension and pressure), which needs to be cleared manually; press \"Peak \" key again, \"AUTO
PEAK \" display, it is the automatic peak state, the displayed test The value is the MAX value in the test
(regardless of tension and pressure), and it will automatically disappear after returning to zero for 2 seconds,
and the next test can be performed.
0002 Push-pull peak mode: the function of simultaneously grasping the MAX load value in both directions of
pressure and tension. When testing the connector, grasp the function of the MAX load on both the positive and
negative sides of the insertion force and the extraction force.
0003 Peak thrust mode-only the MAX load function of insertion pressure is grabbed during the plug-in test
0004 Tension peak mode-when plugging and unplugging, only the function of MAX load that pulls out the
tension is captured.
0005 0006 Switch contact on-off force test mode-accurately measure the load value at the moment of contact on
and off action.
0005 The maximum force value of the external node from the moment of disconnection to the moment of
connection. Connect the two tested contacts to pins 4 and 5 of the data interface, press \"Peak \" to select the
peak mode, and apply pressure to the switch through the push-pull force gauge until the switch is turned on. The
measured force value at this time is the force value required to turn on the switch.
0006 The MAX value of the external node from on to off. Connect the two tested contacts to pins 4 and 5 of the
data interface (using the data plug in the accessory), press the\"\" peak\"key, select the peak mode, and apply
pressure to the switch through the push-pull force gauge until When the switch is open, the measured force
value at this time is the force value required to open the switch.
b. Various setting items