Aliyiqi HF-500 Quick Start Manual - Page 10
Browse online or download pdf Quick Start Manual for Test Equipment Aliyiqi HF-500. Aliyiqi HF-500 15 pages. Hf digital display push-pull force gauge
Press \"Settings \" key in the boot state, the display is \"TEST \", then press \"
options. \" HIDT \", \" LODT \", \" STOP \", \" LOSET \", \" OFFT \", \" PE.SET \", \"GSET \", \"BACSET \",
\"RESET \". As shown below:
(HIDT) Test value upper limit setting: Set the test value upper limit, the upper limit default state is
99% of full scale. Above the upper limit is out of range, \"MAX \" is displayed. If you want to reset the upper limit,
use \"
\"Key and \"
(LODT) Test value lower limit setting: Set the test value lower limit, the default value is 0 by default.
Below the lower limit is out of range, \"MIN \" is displayed. If you want to reset the lower limit, it is similar to the
previous operation, use \"
(STOP) The machine stop value setting, the default state is 99% of full scale. This product is used in
combination with our own motor table. When the test value exceeds the set machine stop value, the instrument
outputs a 5V signal and the machine receives the signal.
Operation method of this function:
\"Key setting. \" Save \"and return to the setting item interface. As shown below:
\"Key and \"
\"Key setting. \" Save \"and return to the setting item interface.
\"Key, you can view the setting