PG7000 ™ Upgrade Instructions Flash with EEPROM to Flash without EEPROM,
Flash the PG7000 with PG7000 software ver. 2.02d. (This is available for download at DHI web site:
www.dhinstruments.com). Refer to the instructions provided in Section 5.5 of the PG7000 manual. Remember, a
"null modem" cable must be used to connect the PC to the PG7000 COM2 port.
Reboot the PG7000 following the reboot instructions displayed. When asked whether to upgrade answer, "YES".
Repeat step
and
PG7000 software upgrades are designed to reformat and transfer data as needed. However, various PG7000 user
settings may be reset and/or deleted. The extent depends on the number of revisions between the original and the
newly flashed version. Be sure to check the metrological aspects including piston-cylinder, mass set, mass bell and
internal sensor calibration files as these may be completely deleted and/or reset. See the PG7000 manual.
Remove EEPROMs from U41 and U46.
Deliver the upgraded PG7000 Platform to the customer with the new PG Terminal P/N 401284 and the new PG7000
manual, P/N 550099.
Document 560037c
DH Instruments, Inc.
4765 East Beautiful Lane
Phoenix AZ 85044-5318
USA
with the most recent PG7000 software version.