Daily Set Up & Test Procedure - Ventilator
Close down the bellows to zero
Set the tidal
volume to 600ml
Adjust tidal
volume lock as
required
Attach a one litre reservoir bag to circuit.
Squeeze the self-inflating bag to
demonstrate expansion of test lung (A)
and simultaneous movements of the
reservoir bag on the panel (B).
Set rate to 10
breaths per minute.
Switch on the ventilator,
confirm it works.
B
A
4