Dickey-John GAC 2500 UGMA Manual - Page 9

Browse online or download pdf Manual for Farm Equipment Dickey-John GAC 2500 UGMA. Dickey-John GAC 2500 UGMA 12 pages. Grain analysis computer

ADDENDUM (GAC® 2500 UGMA/INTL)
GAC® 2500 Models (UGMA & INTL)
11001-1646A-201409
To Clean the IR Temperature Sensor:
1. Gently place instrument upside down.
Figure 18
Place Instrument Upside Down
2. Visually inspect the IR temperature sensor to determine the type of
sensor installed.
3. For instruments with the flush sensor and brush bracket, the following
procedure can be utilized to remove particles from the brush. Carefully
place the head of the screw driver (10" length recommended) at the
bristles and use a sweeping side-to-side motion through the entire
length of the bristles no fewer than three times.
Figure 19
Cleaning the Brush Assembly
4. Select the appropriate cotton swab as described. Wet one end of the
cotton swab with 99% Isopropyl alcohol.
MAINTENANCE / 9