Olympus BH2 Series Troubleshooting & Adjustment Manual - Page 39

Browse online or download pdf Troubleshooting & Adjustment Manual for Microscope Olympus BH2 Series. Olympus BH2 Series 46 pages. Phase contrast
Also for Olympus BH2 Series: Complete Teardown, Cleaning, And Reassembly (23 pages), Complete Teardown, Cleaning, And Reassembly (40 pages), Teardown, Cleaning And Reassembly (45 pages), Manual (6 pages), Installation Manual (9 pages)

Olympus BH2 Series Troubleshooting & Adjustment Manual
D.
TROUBLESHOOT-ING '.'(B2..:..BDS-3)
·
No. D-8
4.
Power Supply
4-1
Power is not supplied even after the main switch is turned on,
SIOI
U I 0
Causes
1.
The breaker c:;BlOl functions.
2.
Defective main switch SlOl
3.
Defective AC cord
:
(UYCP15 in lOOV channel)
4.
Defective AC inlet PlOl
5.
Defective breaker CB101
6.
Defective filter PCB Ul04
7.
Defective primary PCB UlOl
Is the breaker functioning?
NO
0 -
--
------,
YES
Press the breaker button.
Is power supplied normally?
; !
YES
NO
.
'