Olympus BH-2 Microscope Motor Drive
Installation
Motor Drive Installation
This procedure steps you through the installation and alignment of the ASI motor drive onto the
Olympus BH-2 microscope. The following tools are required for this procedure:
Small knife or screwdriver
2.5 mm Allen wrench ( Provided )
3.0 mm Allen wrench ( Provided )
1.5 mm Allen wrench ( Provided )
3/32 mm Allen wrench ( Provided )
5/64 mm Allen wrench ( Provided )
7/64 mm Allen wrench ( Provided )
9/64 mm Allen wrench ( Provided )
1/16 mm Allen wrench ( Provided )
Preparing The Microscope
In this part of the procedure the left fine focus knob will be removed from the microscope:
Prior to removing left fine focus knob, the microscope stage should be fully lowered. To do this,
release the lock by rotating fully counter-clockwise the lock ring behind the left coarse focus knob,
then rotate the coarse knob to lower the stage all the way down.
Applied Scientific Instrumentation - https://asiimaging.com/docs/
1/9
Note: the terms left and right refer to the sides of the
microscope as viewed from the front.