Figure 13 – Remove the pivot-adjustment screw

Remove the Center-Pivot Bearing Ball

Use a tweezers to remove the ¼" bearing ball from the
center bore in the stationary base (see
Figure 14 – Remove the center-pivot bearing ball

Remove the Mechanical Detent Stop

The next step is to remove the mechanical detent stop.
Use a suitable JIS screwdriver to remove the two M2X4
pan-head screws securing this stop to the stationary
base of the turret assembly (see
screws are staked in place with adhesive and can be
surprisingly stubborn to remove, so be sure to use the
proper JIS screwdriver to prevent damaging the screw
heads. It might also be helpful to heat the screws with
a heat gun before loosening them, but do not melt the
black knurled-rubber grip ring in the process.
Complete Teardown, Cleaning, and Reassembly of the Olympus BH2-6RE Modular Revolving Nosepiece
Figure
14).
Figure
15). These
Figure 15 – Remove the screws securing the detent stop
Remove the loose mechanical detent stop from the
stationary base (see
Figure
Figure 16 – Remove the mechanical detent stop

Remove the Knurled-Rubber Grip Ring

Carefully remove the black knurled-rubber grip ring
from the outer perimeter of the revolving turret (see
Figure
17), being careful not to stretch or damage it in
the process. This ring is held onto the turret with a few
spots of adhesive, so proceed carefully here.
Figure 17 – Remove the knurled-rubber grip ring
16).