Olympus BH2-CH Substage Assembly - Page 18
Browse online or download pdf Assembly for Microscope Olympus BH2-CH Substage. Olympus BH2-CH Substage 26 pages.
pulling it straight out, away from the main housing (see
Figure 64
and
Figure
62). Reinsert the alignment pins
back into the dovetail block if necessary.
Figure 64 – Remove left dovetail block from the housing
Use a cotton swab and a suitable solvent (e.g., acetone)
to clean the grease from the two dovetail blocks (see
Figure
65).
Figure 65 – Clean grease from the dovetail blocks
Remove the Condenser-Height Knob
Use a 1.5mm Allen wrench or driver to loosen the two
M3x4 set screws securing the condenser-height knob to
the condenser-height shaft (see
Figure 66 – Loosen set screws in condenser-height knob
Remove
the
condenser-height
condenser-height shaft (see
Complete Teardown, Cleaning/Lubrication, and Reassembly of the Olympus BH2-CH Substage Assembly
Figure
66).
knob
from
Figure
67).
Figure 67 – Remove the condenser-height knob
Remove the Condenser-Height Shaft
Use a suitable tool (e.g., sharp needle-nose pliers) to
loosen the lock nut for the tension screw for the
condenser-height shaft (see
Figure 68 – Loosen the tension-screw lock nut
Unscrew and remove the lock nut from the tension
screw (see
Figure
Figure 69 – Unscrew and remove tension-screw lock nut
Use a suitable slotted screwdriver to loosen the
the
condenser-height tension screw (see
Figure
68).
69).
Figure
Revision 1
70).
Page 18 of 26