Olympus BX43 Instructions Manual - Page 24
Browse online or download pdf Instructions Manual for Microscope Olympus BX43. Olympus BX43 48 pages. System microscope
1
20
Fig. 28
1
Fig. 29
3
2
Fig. 30
4
Rotating the Stage
1. Slightly loosen the stage clamping screw @.
2. The stage can be rotated both clockwise and counterclockwise by the
stage clamping screw.
A click may be heard and felt during rotation. However, this is
CAUTION
due to the construction of the substage and does not indicate
a malfunction.
} The angle of rotation varies depending on the positions of the X- and
Y-axis knobs.
Right hand knobs
Left hand knobs
5
Adjusting the Stage Height
} By lowering the position of the substage, the microscope will accommo-
date specimens with maximum height of 35 mm. This is useful when
observing metallurgical specimens and other thick objects.
1. Lower the stage to the lower limit, then remove the stage from the
microscope.
2. Using the Allen screwdriver, loosen the substage bracket clamping screw
@ and remove the substage.
3. Turn the coarse adjustment knob and raise the focusing block 3 to where
the stopper screw 2 on the arm becomes visible.
4. Using the Allen screwdriver, loosen and remove the upper stopper screw
2.
5. Reattach substage bracket and stage.
} Store the removed stopper screw 2 in a safe place so that it will not be
lost, if needed again.
Angle of Rotation
Clockwise
Counterclockwise
230°
20°
20°
230°
(Figs. 29 & 30)
(Fig. 28)