3-2 Focusing Block
Fig. 12
³
²
@
Fig. 13
@
²
Fig. 14
#If the substage is not installed in the correct position, it
may deviate from the coverage range of the damper spring
of the objective and damage the specimen.
²
Ensure that the top surface @ of the substage is flush with the bottom
surface of the notch ² on the mount dovetail. If they are not flush, loosen
@
the clamping screw ³ and lower the substage until it is stopped.
³
1
Replacing the Focus Adjustment Knob
|
# The focus adjustment knob is installed on the right side of the micro-
scope when it left the factory.
}The focus adjustment knob is designed detachable to prevent interfer-
ence with hand during manipulation of the focus adjustment knob or X/
Y-axis knobs.
Usually attach the knob on the opposite side to the stage feed knobs.
1. Loosen the clamping screw @ and remove the focus adjustment knob
².
2. Remove the seal from the focus adjustment knob screw hole on the
other side and attach the knob by reversing the removal procedure.
3. Attach a provided seal on the screw hole | of the focus adjustment dial
³, from which the focus adjustment knob has been removed.
}The focus adjustment dial ³ can be operated with your fingertip or finger
surface at the same time as manipulating the X/Y-axis knobs.
2
F/C button
}This button switches the movement of the stage using the focusing
knob @ or dial between F (Fine) and C (Coarse). (For the purpose of
safety, this button is set automatically to F at the moment the main
switch of the BX-UCB control box is set to " I " (ON).)
· Each press of the F/C button ² switches F and C alternately.
Stage Fine movement:
0.1 mm per turn
Stage Coarse movement: 1 mm per turn
BX61
(Fig. 13)
(Fig. 14)
11