Olympus SZX10 Instructions Manual - Page 25

Browse online or download pdf Instructions Manual for Microscope Olympus SZX10. Olympus SZX10 40 pages. Research high-class stereo microscope
Also for Olympus SZX10: Overview (24 pages), Instructions Manual (36 pages)

Olympus SZX10 Instructions Manual
c
a
b
b
a
e
Downloaded from
ManualsNet.com
a
b
d
c
search engine
3
Mounting the Microscope Body
1
Remove the cap a on the focusing assembly by inserting a thin
object into the notch.
2
Using the Allen wrench, loosen the dovetail mount clamping screw
inside the cap on the focusing assembly by rotating it by 2 or 3 turns
(counterclockwise).
3
Gently insert the dovetail mount c on the rear of the microscope
body into the dovetail mounting port b on the focusing assembly.
Do not insert the mount at an angle or with excessive
NOTE
force, for this may cause malfunctions.
4
When the microscope body has been inserted until it stops, tighten
the clamping screw using the Allen wrench.
5
Place the cap a in the original position.
4
Mounting the Objective
Since the objective is heavy, take the following measures
NOTE
to prepare for an accident in case it is dropped.
· Attach the cap to the objective tip.
· Place the cap of the objective mount thread or a
notebook, etc. on the stand to absorb the shock even
when the objective drops.
While holding the tip of the objective b , mount it on the objective
mount thread a by rotating the objective in the direction of the arrow.
5
Mounting the Observation Tube
1
Using the Allen screwdriver, loosen the observation tube clamping
screw a completely.
2
Aligning the positioning pin c of the microscope body to the
positioning groove b on the observation tube, insert the dovetail
mount d at the bottom of the observation tube into the mounting
port e on the microscope body.
3
Using the Allen screwdriver, tighten the observation tube clamping
screw a .
The observation tube can be mounted 180° from the
NOTE
above position, but this positioning makes observation
difficult and is not recommended.
When the auxiliary pillar is used, this positioning is
impossible because the eyepiece gets in the way.
SZX10
21