Troubleshooting
This section describes the possible problems of this product and their remedies.
If problems occur, please review the following list and take remedial action as needed.
If you cannot solve the problem after checking the entire list, contact Olympus for assistance.
Phenomena
Even though the power switch is
set to (ON), the pilot lamp does
not turn ON.
The pilot lamp does not turn ON by
pressing the lamp switch.
The lamp is too bright.
The pilot lamp is blue, but the light
from the lamp does not illuminate
the sample.
The lamp is too dark.
The lamp does not turn ON.
The U-LGPS control software in use
stopped.
The liquid light guide cannot be
attached to U-LGPS or U-LLGAD.
Repair request
If you cannot solve the problems even though taking actions described in Troubleshooting, please contact Olympus
for assistance. At that time, please tell them the following information as well.
· Product name and abbreviation (Example: U-LGPS)
· Serial Number
· Phenomena or strings displayed on the counter on the front of U-LGPS.
Cause
The power cord is not connected
properly.
The power cord plug is not
connected properly.
The fuse is blown out.
The power switch is set to
–
The shutter is in the light path of
the microscope frame.
The liquid light guide is not
attached to the U-LLGAD properly.
Or U-LLGAD is not attached to the
illuminator of the microscope frame
properly.
The pilot lamp is green.
–
The liquid light guide other than
the conformable liquid light guide
is used.
Remedy
Connect the power cord properly.
Connect the power cord plug
properly.
Replace the fuse with a new one.
(OFF). Set the power switch to (ON).
Insert the optical element in the
light path of the microscope frame
to adjust the brightness.
For details, refer to the instruction
manual for the microscope frame
you use.
Remove the shutter from the light
path.
For details, refer to the instruction
manual for the microscope frame
you use.
Connect them properly.
Press the lamp switch to make the
pilot lamp blue.
Set the power switch to
Then, turn ON the lamp.
Use the conformable liquid light
guide.
U-LGPS
Page
16
16
17
11
-
-
15
12
(OFF).
11
14
13