Olympus PSD-10 Instructions Manual - Page 6

Browse online or download pdf Instructions Manual for Medical Equipment Olympus PSD-10. Olympus PSD-10 28 pages. Electrosurgical unit

Output check
'Monitors
Connection monitor
Feedback ratio monitor
S-P cord
A-cord
Connection
Check lead
Footswitch
Type of protection
against electric shock
Classification
Degree of protection
(Electro-
against electric shock
medical
equipment)
Degree of protection
against explosion
Other
Non-Controlled Document
• When the Check Lead is connected and the footswitch
depressed, either the "OK" or "FAULT" light will come
on.
• Checks the PSD-10 or the entire electrosurgical system.
The entire electrosurgical system cannot be tested when
using PSD-10 in combination with CD-3U/L or CD-40/L.
Detects defective connections of cords and broken wires
in the S-P cord, and activates:
• A warning light (P-cord or S-cord symbol), a warning
buzzer, and inhibits output.
Computes return current/delivered current ratio. If the
ratio is less than a preset value, it inhibits output.
Plug-in (PSD-10) and screw-on (P-plate and endoscope)
Plug-in
Plug-in
Screw-on
Class I equipment
Type BF (Body Floating) equipment
The PSD-10 should never be used where there is a risk
of flammable gases.
Defibrillator-protected equipment