FEATURES AND MAIN SPECIFICATIONS
1-1
Features
1. Simplified pre-procedural test, using a Check Lead, of the power output of
the PSD-10 alone as well as that of the entire electrosurgical system includ-
ing the endoscope and endoscopic electrodes. Test results confirmed by
TEST "OK" or "FAULT" indicator lights on the front panel.
2. Push-button selection of CUT, COAG or BLEND current. Output power is
controlled by a single-pedal footswitch.
3. The connection monitor circuit detects loose or inadvertent misconnection
of cords or broken wires in the S-P cord (Safety cord and Patient cord).
If any irregularity is present, the monitor activates an alarm and inhibits
output (current is not delivered when the footswitch is depressed) to ensure
safety of the procedure.
4. The feedback ratio monitor scans the entire circuit when the footswitch is
depressed. If excessive leakage current is present, the output is inhibited to
ensure safety of the procedure.
5. Flat switches and display panel can be easily wiped clean.
6. Compact, lightweight and portable.