Olympus BH-NRE Manual - Page 13

Browse online or download pdf Manual for Laboratory Equipment Olympus BH-NRE. Olympus BH-NRE 17 pages. Modular revolving nosepiece

Olympus BH-NRE Manual
dovetail slide with the four corresponding tapped holes
in the stationary base (see
the dovetail slide such that the relief notch (shown in
the inset of
Figure
37) is facing downwards and towards
the slotted lock ring and pivot-adjustment screw in the
center of the stationary base.
Figure 37 – Position dovetail slide onto stationary base
While holding the dovetail slide in this position on the
stationary base, use a suitable JIS screwdriver to
reinstall four M2.6X5 pan-head screws to secure the
dovetail slide onto the stationary base of the turret
assembly (see
Figure
38).
Figure 38 – Secure dovetail slide onto stationary base

Apply Grease to the Mechanical Detents

Apply fresh grease (item S1 of
mechanical detent notches in the revolving turret (see
Figure
39).
Be careful that you do not get grease
anywhere else.
Complete Teardown, Cleaning, and Reassembly of the Olympus BH-NRE Modular Revolving Nosepiece
Figure
37). Be sure to orient
Appendix
2) to the four
Figure 39 – Apply grease to the four detent notches

Reinstall the Protective Cover

Place the protective cover into position on the back side
of the turret assembly, lining up the three holes in the
protective cover with the three tapped holes in the
stationary base (see
Figure
Figure 40 – Place cover in position on the turret assembly
Use a suitable JIS screwdriver to reinstall three M2X3
countersink screws to secure the protective cover onto
the stationary base (see
Figure 41 – Secure protective cover using three screws
40).
Figure
41).
Page 13 of 17
Revision 3