Olympus CIX100 Instructions Manual - Page 25

Browse online or download pdf Instructions Manual for Laboratory Equipment Olympus CIX100. Olympus CIX100 46 pages. Cleanliness inspector system
Also for Olympus CIX100: Systems Setup Manual (38 pages), Installation Manual (14 pages)

Olympus CIX100 Instructions Manual
If the white balance is not yet calibrated for all sample types and
l
observation methods, [White Balance] will be marked with a yellow
exclamation mark.
2. Select [White balance] and click [Calibrate].
In the following dialog box, you find a list of observation methods. In
l
your CIX100 system, each observation method reflects a list of
optimized software settings for the analysis under specific optical
conditions. Each sample type requires its own observation method. If
your system is equipped with a real color slider, two observation
methods will have to be calibrated for each sample type for optimal
results. The observation method with real color slider is named after
the particle type appended with "in Real Color Mode" and is always
listed below the observation method for the respective particle type
without real color slider.
Sample Holders
25