U-EXCITATION FILTER SET CKX-NU (OPTIONAL)
}The CKX-NU U-Excitation Filter Set should be mounted by the dealer in the position (fluorescent mirror
slider position) specified by the customer. (It is also possible, on custom order, to replace the U-excita-
tion filters with excitation filter, barrier filter and/or dichroic mirror of the UIS fluorescent mirror unit series.)
Contents of the set
Excitation filter
Construction of the fluorescent mirror slider
Mirror holder
Removing the UV cut filter (Fig. 25)
Loosen the holder ring 2 of the UV cut filter 1 in the front of the collector of the U-LH50HG lamp
housing by turning the ring counterclockwise, and then take out the filter.
29
Barrier filter
Filter holders (x 2)
Filter holder
Barrier filter
Dichroic mirror
Indication stickers
(To be attached on the
light path indicators)
Excitation filter
All pieces can be placed and replaced using
a precision screwdriver.
Mirror holder
Filter holders
BF light path shield plate
1
2
Fig. 25