Olympus MX61A Manual - Page 30

Browse online or download pdf Manual for Laboratory Equipment Olympus MX61A. Olympus MX61A 48 pages. Motorized 300 mm compatible semiconductor/fpd inspection microscope

USER screen 1
@
|
USER screen 2
PASSWORD screen
27
1

Initialization

}Be sure to set DIP SW1 bit 7 of the BX-UCB control box to "ON".
}Do not alter the settings of the DIP switches of the Z-board and AF board.
1. Set the main switches of the BX-UCB control box and MX-OPU61A op-
eration unit to " " (ON).
The microscope's Z-stage will be initialized and the application will start
up automatically.
}It takes about 40 seconds to start up the application.
}The initialization time of the Z-stage varies depending on the position of
the Z-stage. The startup of the application will not finish during the
initialization of the stage (an error message will be displayed).
2. When the MX-OPU61A is used for the first time, message "Please restart
BX-UCB" is displayed on the screen. When this message is displayed,
set the main switch of the BX-UCB to "
3. When the BX-UCB is turned ON again, the application installed in the
MX-OPU61A starts up automatically. ("INITIALIZING - - -" is displayed until
the USER screen is displayed.)
4. Advance to the next USER screen setup.
}An error message and remedy information are displayed in case of a
trouble. If this happens, follow the instruction of the displayed message.
²
2

USER Screen Setup

}USER switches 1, 2 and 3 can be used to assign three sets of setups for
various observation methods and/or specimens.
1. Touch the USER1 (or 2 or 3) switch @.
If this is the first time, the default settings are displayed because no CONFIG
information has been set up. (An alarm that the actual unit information
differs from the setting information is displayed at this time.)
ƒ
2. Touching one of the USER switches enables the SET UP switch ² that
has previously been grayed out. Touch the SET UP switch to display the
PASSWORD screen.
3. Input "69612" as the password. (The password is fixed.)
4. Touch the OK switch ³ to display the CONFIG 1/6 screen.
}An alarm message is displayed if the input password is incorrect.
# Set the upper and lower limits of the focusing mechanism of CONFIG
2/6 every time the thickness of a specimen changes.
Other Switches
|UNIT INFO:
Touch this switch to open the UNIT screen, in which the unit connection
status after initialization is displayed.
³
Revolving nosepiece connection status ---- > REVO:
DIC prism unit connection status ---- > DIC:
Analyzer unit connection status ---- > AN:
Hand switch connection status ---- > HS:
AF unit connection status
ƒOK:
Touch this switch to display the observation operation screen.
" (OFF) then " " (ON) again.
"5" with a 5-position revolving nosepiece.
"6" with a 6-position revolving nosepiece
"CONNECT" when connected,
"- - -" when not connected.
---- > AF: