Olympus MX61A Manual - Page 32

Browse online or download pdf Manual for Laboratory Equipment Olympus MX61A. Olympus MX61A 48 pages. Motorized 300 mm compatible semiconductor/fpd inspection microscope

@
CONFIG 1/6 screen
²
|
AS preset selection screen
29
Setting the AS preset values (AS PRESET)
}After the objective names and magnifications have been set up, the rec-
ommended AS preset values are set automatically. The preset values
extend in the range from 1 to 13 (aperture iris diaphragm fully open to
fully closed). If "- - -" is set, the interlocking is disabled.
1. If it is required to modify a recommended preset value, touch the
switch @ below the preset value to open the AS preset selection screen.
2. Touch the desired AS preset value ² and touch the OK switch ³ to enter it.
If the desired AS preset value is not shown in the AS preset value selec-
tion screen, touch the page up/down switch | until the desired AS pre-
set value is displayed (a total of 3 pages are provided).
3. Touch the OK switch to display the CONFIG 1/6 screen again.
AS Preset Values
OB magnification
OB type
UMPlan FL
UMPlan FL-BD
³
MPlan FL
MPlan FL-BD
LMPlan FL
LMPlan FL-BD
MPlan FLN
MPlan FLN-BD
LMPlan FLN
LMPlan FLN-BD
MPlan N
MPlan N-BD
LMPlan Apo
LMPlan Apo-BD
1.25 2.5
5
10
20
--
--
3
3
4
--
--
3
3
4
--
--
4
4
5
2
2
3
3
4
--
--
--
--
4
4
5
--
--
3
3
4
--
--
--
--
--
40
50 100 150
5
6
8
--
--
--
6
8
--
--
8
9
--
--
--
6
8
9
--
8
9
--
--
6
8
--
--
--
--
10