Setting the AS preset values (AS PRESET)
}After the objective names and magnifications have been set up, the rec-
ommended AS preset values are set automatically. The preset values
extend in the range from 1 to 13 (aperture iris diaphragm fully open to
fully closed). If "- - -" is set, the interlocking is disabled.
1. If it is required to modify a recommended preset value, touch the
switch @ below the preset value to open the AS preset selection screen.
2. Touch the desired AS preset value ² and touch the OK switch ³ to enter it.
If the desired AS preset value is not shown in the AS preset value selec-
tion screen, touch the page up/down switch | until the desired AS pre-
set value is displayed (a total of 3 pages are provided).
3. Touch the OK switch to display the CONFIG 1/6 screen again.