Setting the Option Cube (OP CUBE)
}An option cube (mirror unit) is necessary.
(The default setting for this item is "Nothing".)
· When a DIC prism unit is used, "DIC/Nothing" can be selected.
· When no DIC prism unit is used, "Nothing/PO/FL" can be selected.
1. If it is required to select the option cube, touch the
the OP cube name to open the option cube selection screen.
2. Touch the desired cube name ² and touch the OK switch ³ to enter it.
3. When the option cube is entered with the OK switch, the CONFIG 1/6
screen is displayed again.
}If "Nothing" is selected, the OP CUBE switches in the operation screens
are grayed out. (The OP light path cannot be selected for observation.)
Other Switches
@FLASHPREV:
This switch turns ON/OFF the function for preventing glare during switch-
ing of the objective in darkfield observation. (The default is "ON".)
²ESC (OB):
This switch turns ON/OFF the function for letting the stage escape dur-
ing switching of the objective. (The default is "ON".)
³CANCEL:
Touch this switch to return to the USER screen without applying the
setups made until then in the CONFIG 1/6 screen.
|NEXT:
Touch this switch to advance to the CONFIG 2/6 screen.
switch @ below