EXTERNAL CONTROL OF REVOLVING NOSEPIECE

1

Connector Used

DX10G1M-14SE (Hirose Electric)
Compatible plug
DX30AM-14P
2

Pin Layout and Signal Names

All of the input/output signals are at the TTL level. The pin layout is shown on each connector.
No.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
3

Revolving Nosepiece Position Numbers (Index Nos. on the Center of Revolving Nosepiece)

Table 2 shows the relationship between the revolving nosepiece position numbers and the STY signals.
Table 1 Signal Names
Signal Name
I/O
N.C.
GND
N.C.
+5V
Reserved
O
OB1
O
OB2
O
OB3
O
N.C.
N.C.
N.C.
N.C.
N.C.
N.C.
Table 2 Revolving Nosepiece Position Numbers
Position No.
OB3
1
2
3
4
5
During rotation
Description
Grounding
+5 V
Reserved for manufacturer.
OB2
1
1
1
0
1
0
0
1
0
1
1
1
– : Negative logic.
OB1
0
1
0
1
0
1
MX61A
40