Olympus SZX-RFL2 Instructions Manual - Page 2

Browse online or download pdf Instructions Manual for Laboratory Equipment Olympus SZX-RFL2. Olympus SZX-RFL2 22 pages. Coaxial fluorescence attachment

Olympus SZX-RFL2 Instructions Manual
IMPORTANT
This equipment can provide the highest level of brightness required in macro fluorescence observation when
combined with the SZX-ZB12 zoom microscope body, DFPLAPO1.2XPF objective and IX-UCLHG/XE collector lens.
# When this equipment is combined with the SZX-ZB9 zoom microscope body, an objective of 0.75X or less
and/or other collector lens than specified above, the brightness and contrast may be degraded.
SAFETY PRECAUTIONS
1. This equipment consists of precision instruments. Handle it with care and avoid subjecting it to sudden or severe impact.
2. The ultrahigh pressure mercury burner used should be the USH102D DC burner (mfd. by Ushio Inc.), that Olympus
supplies.
3. Ensure that the burner is installed and that connection cords are connected properly before operating the unit.
4. Do not open the lamp housing while the lamp is turned on, or for at least 10 minutes after the lamp has been turned off.
Lamp housing parts are extremely hot and will cause burns if touched. (See page 9.)
5. Do not apply excessive force to the stoppers which are provided for some functions. Otherwise the stopper may be
damaged.
6. Do not attempt to open or disassemble the power supply unit because it includes high voltage parts inside.
7. To ensure safety, be sure to ground the power supply unit. Otherwise the intended electrical safety performance cannot be
obtained.
8. Before opening the lamp housing for replacement of the burner or any other internal part, set the main switch to " " (OFF),
unplug the lamp housing's connecting cord plug from the output connector on the power supply unit, and wait for 10
minutes or more, until the lamp housing cools down.
9. Before replacing the mirror unit, be always sure to set the shutter slider of the illuminator tube in the fully pulled-out (light-
excluding) position or set the main switch of the power supply unit to "
Safety Symbols
The following symbols are found on the equipment. Study the meaning of the symbols, and always use the equipment
in the safest possible manner.
Symbol
Indicates the presence of high voltage (1 kV or more). Take caution to guard against electric
shock.
Indicates that the surface becomes hot, and should not be touched with bare hands
Before use, carefully read the instruction manual. Improper use could result in personal injury
to the user and/or damage to the equipment.
l
Indicates that the main switch is ON.
\
Indicates that the main switch is OFF.
Warning Indications
Warning indications are placed at parts where special precaution is required when handling and using the equipment.
Always heed the warnings.
Warning indication
position
Lamp housing (U-ULH)
Burner socket (U-ULS100HG)
Lamp housing (U-ULH)
i
" (OFF).
Explanation
[Warning against high temperature]
[Warning against high voltage]