REFILL TEST NUMBER
-
Press the SETTING button
-
Select the REFILL item
-
Insert the TEST DEVICE Original 4K in the dedicated
compartment (as shown in the figure)
-
Press the CHECK button
-
Press OK to confirm the number of tests to refill
-
Automatic update of the number of tests displayed on
the instrument screen (top left)
TROUBLESHOOTING
ERROR
MESSAGES
Low (Lo)
High (HI)
ERR
Exam not
available.
Need to
reload
USB Pendrive
not found
Attention!
"n" test remain
TEST DEVICE
Already used
Rev. 0 – 30/09/2022
CAUSE
Blood level inside the test
tube too low
Blood level inside the test
tube too high
Interference caused
by labels
Generic error, pressing the
error message will display a
number that differentiates
the cause. See the complete
TROUBLESHOOTING
in the User Manual
The executable test
counter is equal to zero
The instrument does not
recognize the USB Pendrive
The number of tests
remaining is less than 50
The used Test Device is
empty
HELP
If possible, add blood up to
established levels
If possible, remove blood up to
established levels
Try to remove the label and
repeat the test
Verify that the tube used is
compatible with the instrument.
If there are more than 2 labels,
remove the excess labels.
Check for any lumps or clots.
Try to repeat the test
Follow the instructions described
at the top of this page to recharge
the instrument with a new Test
Device
Remove and reinsert the device
or replace it
It is recommended to recharge
the instrument as soon as
possible with a new Test Device
Repeat with a new Test Device