Replacement of the 15" LCD Color TFT Display
Geometry
Requirements
•
The basic unit and the monitor cart are connected to each other;
the system is fully functional.
Fig. 4
•
The basic unit and monitor cart are connected to each other. The Memoskop is
synchronized by the VIDEOMED DC camera.
•
At the MEMOSKOP, call up the SMPTE test pattern. (Technical setup)
•
Check the geometry of the displayed test pattern. It should be displayed as shown
in Fig. 4. If adjustments are necessary, follow the description "Manual programming"
(page 16).
•
Disconnect the basic unit from the monitor cart (disconnect X10). The MEMOSKOP is in
stand alone mode (internal synchronization).
•
Check the geometry of the displayed test pattern. It should be displayed as shown
in Fig. 4. If adjustments are necessary, follow the description "Manual programming"
(page 16).
Siremobil Compact L
SPR2-130.841.03
Rev. 02
06.05