Siemens SIREMOBIL Compact L Instructions Manual
Replacement of the 15" LCD Color TFT Display
Replacing the switched-mode power supply
NOTE
Fig. 2
Disconnect the low voltage plug from the TFT display.
Disconnect the power plug of the switched-mode power supply.
Loosen the cable ties of the low voltage cable and pull it back to the log book
compartment.
Take out the switched-mode power supply from the log book compartment.
Install the new switched-mode power supply in the log book compartment, using new
cable ties.
Install the low voltage cable to the TFT display and connect it.
Fix the low voltage cable with cable ties.
Connect the power plug to the switched-mode power supply.
Check the function of the TFT display.
Siemens AG
Medical Solutions
The switched-mode power supplies are installed in the log book
compartment and are fixed with cable ties.