Replacement of the 15" LCD Color TFT Display
Replacing the switched-mode power supply
NOTE
Fig. 2
•
Disconnect the low voltage plug from the TFT display.
•
Disconnect the power plug of the switched-mode power supply.
•
Loosen the cable ties of the low voltage cable and pull it back to the log book
compartment.
•
Take out the switched-mode power supply from the log book compartment.
•
Install the new switched-mode power supply in the log book compartment, using new
cable ties.
•
Install the low voltage cable to the TFT display and connect it.
•
Fix the low voltage cable with cable ties.
•
Connect the power plug to the switched-mode power supply.
•
Check the function of the TFT display.
Siemens AG
Medical Solutions
The switched-mode power supplies are installed in the log book
compartment and are fixed with cable ties.
SPR2-130.841.03
Rev. 02
Page 9 of 20
06.05
CS PS 24
9
1
Siremobil Compact L