Replacing the horizontal lift brake
2
Fig. 16
Fig. 18
Installation
•
Lift the horizontal brake out of the bearing support. Reserve the two springs (2/Fig.16).
•
Remove both screws from the old horizontal brake (1/Fig. 16) and install them on the
new horizontal brake.
•
Place the new horizontal brake in the bearing support. The two screws should not stick
in the slots. (Check this by pressing down on the horizontal brake). You may have to
take out the horizontal brake again and slightly bend the screws to the side (hit the screw
head lightly on the side)
•
Route the horizontal brake cable back through the plate (1/Fig.17). The cable must be
positioned in the recess (1/Fig.18, 1/Fig.19).
•
Route the cable end of the horizontal brake parallel to the cable bundle in the horizontal
carriage.