4
If a multiformat camera is not connected
Power cable
Videoprinter with 100VAC line voltage:
Connect the line wires to transformer T1, connector TI - 34 and TI - 36.
Connect the protective conductor to the protective conductor terminal stripe (ye/gn).
Videoprinter with 120VAC line voltage:
Connect the line wires to transformer T1, connector TI - 35 and TI - 36.
Connect the protective conductor to the protective conductor terminal stripe (ye/gn).
Videoprinter with 230VAC line voltage:
•
Connect the line wires of the video printer at terminals X2.1 and X2.2.
•
Connect the protective conductor at the corresponding location (yellow/green).
•
Route the line wires to the video printer compartment.
Hardcopy exposure control cable
•
Connect the control cable included with the shipment at the Memoskop "Printer"
connector.
•
Route the end of the cable with the 3.5 mm jack plug to the video printer compartment.
Video cable
•
Connect the video cable at the Memoskop "Out3" connector.
•
Route the video cable to the video printer compartment.
•
Connect the video printer. Ensure the proper termination of the video cable
(75 ohm).
Programming
For Memoskop C (with keyboard)
•
Switch on the system.
•
Press [CTRL] and [U] simultaneously on the Memoskop keyboard.
•
Call up the "C Documentation" sub-menu.
•
Perform the configuration on monitor A for SIREMOBIL Iso-C systems with one monitor.
•
Perform the configuration on monitor B for SIREMOBIL Iso-C systems with two monitors.
•
If the system is equipped with an area dose measurement device (Diamentor), call up
the "K documentation of the dose" sub-menu and configure to "yes".
•
End the "User menu".
SIREMOBIL Iso-C
This calls the "User Setup menu" for Memoskop.
Register 3
Installing a Video Printer
SPR2-230.033.02
Rev. 02
10.99
Page 4 of 6
TD SD 24
Medical Engineering
1
1
1
1
1
Siemens AG