After SITRANS WW300 has been properly installed, calibration of the weighing system
must be done in conjunction with the integrator. Refer to the integrator instruction
manual for programming and calibration. The calibration is initially done using the
supplied test load. Material tests are recommended to achieve maximum accuracy.
Test Load
The test load value for your SITRANS WW300 is given on the accompanying factory test
sheet, in kilograms per meter or pounds per foot.
Zero
Perform the zero calibration as described in the Calibration section of the integrator
manual.
Span
The test load used in the calibration procedure is a set of factory sized and supplied test
weights (2 or more).
Perform the span calibration as described in the Calibration section of the integrator
instruction manual.
After the span calibration has been completed, remove the test load and store it.
7ML19985MQ01