SENC 150 Reading Head Replacement Instructions
Preparing the removal
Remove all cable ties or retaining clips.
Disconnect the reading head cable from the readout.
Installing alignment brackets
Attach the alignment brackets (included in head replacement
hardware kit) by inserting the top ears into the bottom grooves
in the scale case.
Numbers and letters on reading head casting and alignment
brackets correspond when assembled.
Slide brackets against
reading head snap into place.
2 Preparing the removal
Insert brackets at 45°
and twist to lock in scale grove.