OPERATING INSTRUCTIONS
Pre-test Conditions
1. Operating temperature range is -10° C to 50° C. Use within this temperature
range will limit condensation from a subject breath to insure an accurate
reading from a breath sample.
2. It is recommended that the operator to observe a subject for about 20 minutes
before administering the breath test to insure that nothing is consumed or
expelled from the mouth of the subject.
eliminate reading from mouth alcohol.
3. To ensure the fuel cell is kept in good condition, turn off the device after use
and store it in appropriate condition.
Standard Test
1. Attach a disposable mouthpiece onto the sample port.
2. Input information by a stylus pen.
NOTE: Maximum 24 characters for each input.
3. When displays "Ready, Please Blow", take a deep breath and blow until
BEEP tone stops.
4. A photo will be taken while the testee is blowing.
5. A passive test can be performed by pressing
blowing.
6. Test result displays on the display.
7. If testee did not blow within 30 seconds, tester can press
testee refused to do the test; or press
again.
8. If testee failed to give a valid breath sample, tester can press
or press
to allow testee to do the test again.
Following this procedure will
during the testee is
to allow testee to do the test
P. 5
to confirm the
to confirm;