Diverse MF500M+ User Manual - Page 6

Browse online or download pdf User Manual for Measuring Instruments Diverse MF500M+. Diverse MF500M+ 16 pages. Magmeter

Operation - Options

This is the normal mode of operation.
There are two display modes for this menu which can be
toggled using a long press on the up � or down � keys. Data
is updated approximately every second.
The first mode displays the measured value, currently stored
offset and signal (difference) on the bottom line. The next
available record number is shown on the top line.
The second display mode estimates the amount of pipe
occlusion based on the current measured and offset values
(note there are a number of assumptions in these estimates see
section on calibration).
Note that data displayed on the first of these display formats is
what is recorded on a short probe button press.
Press tick � to view stored records (4 records shown, 1 record
per line). Tick � toggles between the two display formats.
The default format is record number, measure value, stored
offset and signal (difference between measured and zero/offset).
Alternative format gives record number and signal.
Use the up � or down � keys to scroll through stored
records. Long press of up � or down � scrolls a page i.e. 4
records. Press long tick � to return to <Measure>.
Page 6