Operation - Options
Press tick � to clear records and set next record pointer to
start of memory. Reverts back to main screen.
Press tick � to see software version. To leave press up � or
down � keys.
The instrument has two calibration factors used when
estimating pipe occlusion, gain and packing factor. The
following section gives details on how these factors are
calculated.
This menu option gives the user the opportunity to modify the
calibration factors. Press tick � to see factors.
A long press on the up � or down � keys is used to select
the factor. The current selection is shown by the symbol >.
Press up � and down � keys to increase or decrease the
selected factor. With gain or packing selected press enter to
store the values and exit to the main menu.
To revert to the factory calibration press enter button with
default selected.