Arthur Grillo DPC200-R Installation And Operation Manual - Page 14
Browse online or download pdf Installation And Operation Manual for Controller Arthur Grillo DPC200-R. Arthur Grillo DPC200-R 16 pages. Airflow / pressure controller
WARRANTY & SUPPORT
Measuring mode:
For limit value monitoring a limit value can be entered. This value is set in the menu item
„limit switch". The previously set parameters are taken into account (unit, parameter,
k-factor and the measuring range).
As default setting the limits are not active - Display - 2nd line: „OFF"
Alarm ON: Measurement for set delay time constantly greater than 1.0 • limit value
Alarm OFF: Measurement for set delay time constantly less than 0.95 • limit value
6.
Maintenance
The DPC200-R contains no wearing or consumable parts. Servicing is not required.
On request, Arthur Grillo GmbH offers an annual calibration with factory certificate.
For information, please contact:
Arthur Grillo GmbH
Am Sandbach 7
40878 Ratingen
7.
Warranty
Warranty and liability claims for personal and property damage are excluded if they are
caused by one or more of the following reasons:
•
Improper use of the device.
•
Improper installation, commissioning, operation and maintenance of the device.
•
Unauthorized modifications to the device beyond the intended use.
•
Disasters due to external influences and force majeure.
8.
Troubleshooting
Description
Display does not show
anything
Measurement stays zero
Measuring error
Error remains
Display shows:
14
Phone: +49 21 02 - 47 10 22
Fax:
E-Mail: [email protected]
Activity
Check electrical connection
Function test with a slight pressure increase in measu-
ring mode for differential pressure
Perform zeroing as described in chapter 4.5.
Contact manufacturer
!
The device is in the alarm state, see section 5.7
Doc.-no.: DPC200-R_02_EN
+49 21 02 - 47 58 82
Issue: 06/2018