CUSTOM ULTRASONICS 20000 Operator's Manual - Page 3
Browse online or download pdf Operator's Manual for Storage CUSTOM ULTRASONICS 20000. CUSTOM ULTRASONICS 20000 7 pages. Endoscope storage cabinet
Also for CUSTOM ULTRASONICS 20000: Operator's Manual (11 pages)
CUSTOM ULTRASONICS, INC.
I
:
NSTALLATION
Secure the cabinet to the wall using the brackets located along the top of the rear wall of the cabinet.
When mounting your cabinet to the wall, do not drill holes through the interior walls of the cabinet.
C
: Cabinet will tip over if not properly secured to wall and doors are opened.
AUTION
Insert the hospital grade grounded plug into a 110 VAC, 15 – 20 AMP, 60 Hz, 1 receptacle. The
plug is located on the top of the cabinet.
The upper door of the Cabinet provides access to the desiccant chamber. Place the six (6) bags of
desiccant into this chamber. Place one desiccant card in between two desiccant bags. The door of
the chamber must remain closed during use of the cabinet.
The programmable timer can be adjusted to operate the air pump for the desired amount of time
from one (1) to thirty (30) minutes every hour. Initially, set the timer to operate for ten (10) minutes
per hour, and increase by two (2) minute increments until the desired humidity level, as determined
by the hospital/facility, has been achieved. Humidity indicator cards, placed between the desiccant
bags, can be used to assist in measuring the humidity level of the desiccant. Drier climates and less
frequent opening of the door will reduce pump usage. Conversely, more humid climates and/or
more frequent opening of the cabinet's door will require increasing the pump run time via the timer.
Inside the cabinet on its front side is a gray electrical box. Located on the rear of this box is a
momentary switch. When the momentary switch is engaged, the air pump will function. This switch
is used for testing the operation of the pump.
Activate the momentary switch daily to confirm proper operation of the air pump.
SHEET
REVISION
3 OF 7
A
M
20000 E
ODEL
NDOSCOPE
S
C
TORAGE
ABINET
O
'
M
PERATOR
S
ANUAL